Instruction Manual
KD-2300
Noncontacting Displacement Measuring System
EC DECLARATION OF CONFORMITY
This apparatus, when installed and operated per the manufacturer’s
recommendations, conforms with the protection requirements of EC Council
Directive 89/336/EEC on the approximation of the laws of the member states
relating to Electromagnetic Compatibility, as required by the Technical Construction
File Route to Conformity.
This certificate has been issued in accordance with the conditions of regulation No.
53 of U.K. Statutory Instruments 1992 No: 2372 – The Electromagnetic
Compatibility Regulations.
********
DO NOT MAKE ANY MODIFICATIONS TO CABLE LENGTH, SENSOR
OR CALIBRATED TARGET MATERIALS WITHOUT PRIOR
CONSULTATION WITH A KAMAN APPLICATION ENGINEER
*********
Copyright © 2000
PART NO: 860021
Rev. F
Zobrazit stránku 0
1 2 3 4 5 6 ... 43 44

Shrnutí obsahu

Strany 1 - KD-2300

Instruction Manual KD-2300 Noncontacting Displacement Measuring System EC DECLARATION OF CONFORMITY This apparatus, when installed and

Strany 2 - Contents

KD-2300 Instruction Manual Part I: What is the KD-2300? •••• 7 Part I: What is the KD-2300? System Technology Kaman Precision Products’s KD-2300 fam

Strany 3 - Appendix A: Glossary 33

8 •••• Part I: What is the KD-2300? KD-2300 Instruction Manual Nonconductive materials intervening between the sensor and target have little or no e

Strany 4 - Introduction

KD-2300 Instruction Manual Part II: Getting Started •••• 9 Part II: Getting Started Basic System Your KD-2300 system consists of these basic compon

Strany 5 - Service Information

10 •••• Part II: Getting Started KD-2300 Instruction Manual !Connecting External Power Supply The KD-2300 operates from a ± 15 Vdc power source. I

Strany 6 - Kaman Products Standard

KD-2300 Instruction Manual Part II: Getting Started •••• 11 15 VSupply+15 VSupply-165432DVM,Recorder,Oscilloscope,orComputer Interface+-+ Output- O

Strany 7 - Using this Manual

12 •••• Part II: Getting Started KD-2300 Instruction Manual

Strany 8 - Icon Directory

KD-2300 Instruction Manual Part III: The Equipment •••• 13 Part III: The Equipment Operating Principle* The system uses a principle of impedance var

Strany 9 - Application Information

14 •••• Part III: The Equipment KD-2300 Instruction Manual Accessories Recommended accessories for ease of system calibration and convenience are li

Strany 10 - Part I: What is the KD-2300?

KD-2300 Instruction Manual Part III: The Equipment •••• 15 Sensor Options

Strany 11 - System Configuration

16 •••• Part III: The Equipment KD-2300 Instruction Manual Sensor Specification The table below lists the sensors available for use with the KD-230

Strany 12 - Part II: Getting Started

KD-2300 Instruction Manual Contents •••• i Contents Introduction 1 Welcome...

Strany 13

KD-2300 Instruction Manual Part III: The Equipment •••• 17 Measuring Range The specified linear measuring range for an inductive system is directly

Strany 14 - Figure 1

18 •••• Part III: The Equipment KD-2300 Instruction Manual material (other than the target) that is positioned within 3 times the diameter of the se

Strany 15

KD-2300 Instruction Manual Part III: The Equipment •••• 19 Surface Mounting Surface mounting will not affect a sensor’s calibration assuming there

Strany 16 - Part III: The Equipment

20 •••• Part III: The Equipment KD-2300 Instruction Manual Flush Mounting Flush mounting in a conductive fixture requires a cut-out area around the

Strany 17

KD-2300 Instruction Manual Part III: The Equipment •••• 21 Target Material The essential requirement for a proper target is that it be conductive.

Strany 18 - Sensor Options

22 •••• Part III: The Equipment KD-2300 Instruction Manual under the face of a sensor. The diagram on the right shows the corresponding analog volt

Strany 19 - Sensor Specification

KD-2300 Instruction Manual Part IV: Calibration •••• 23 Part IV: Calibration Factory Calibration All systems are factory calibrated; even so, we rec

Strany 20 - Conductive vs. Nonconductive

24 •••• Part IV: Calibration KD-2300 Instruction Manual Offset Offset is the minimum space or gap between the sensor and target. You should positi

Strany 21

KD-2300 Instruction Manual Part IV: Calibration •••• 25 Calibration Controls You will need to calibrate each measuring channel in your system using

Strany 22 - Surface Mounting

26 •••• Part IV: Calibration KD-2300 Instruction Manual Gain Control Gain control affects the change in output of a system in volts due to a given

Strany 23 - Flush Mounting

Contents •••• ii KD-2300 Instruction Manual Part IV: Calibration 23 Factory Calibration ...

Strany 24 - Target Size

KD-2300 Instruction Manual Part IV: Calibration •••• 27 VDCoffsetDISPLACEMENTFull Scale CalibrationVDCoffsetDISPLACEMENTBipolar Output Calibration

Strany 25 - Other Considerations

28 •••• Part IV: Calibration KD-2300 Instruction Manual 1. Install the sensor in the calibration or application fixture, ensuring that when the ta

Strany 26 - Part IV: Calibration

KD-2300 Instruction Manual Part IV: Calibration •••• 29 VDCoffsetDISPLACEMENTBipolar Output Calibration Use this method when your application is be

Strany 27

30 •••• Part IV: Calibration KD-2300 Instruction Manual 6. Repeat Steps 3 through 5 as many times as necessary until you reach the desired output

Strany 28 - Zero Control

KD-2300 Instruction Manual Part IV: Calibration •••• 31 High Accuracy Band Calibration DISPLACEMENToffsetVDC0High Accuracy BandCalibrationCALIBRATED

Strany 29 - Linearity

32 •••• Part IV: Calibration KD-2300 Instruction Manual 1. Install the sensor in the calibration or application fixture, ensuring that when the ta

Strany 30 - Full Scale Calibration

KD-2300 Instruction Manual Appendix A: Glossary •••• 33 Appendix A: Glossary Glossary A/D Converter A device that converts an analog voltage to a d

Strany 31

34 •••• Appendix A: Glossary KD-2300 Instruction Manual Measurement The difference between the upper and lower frequency response limits of a Ban

Strany 32

KD-2300 Instruction Manual Appendix B: System Modifications •••• 35 Appendix B: System Modifications Maximum Sensitivity (Gain Control) When increas

Strany 33

36 •••• Appendix B: System Modifications KD-2300 Instruction Manual GCut this trace formaximum sensitivity (gain)LHCut this trace for least sensitiv

Strany 34

KD-2300 Instruction Manual Introduction •••• 1 Introduction Welcome Welcome to the Series 2300 by Kaman — a family of precision, noncontact displace

Strany 35 - will improve linearity. In

KD-2300 Instruction Manual Appendix B: System Modifications •••• 37 E1 & E2 inthis areaE3, E4, & E5in this areaLower PC Board FOR SLAVE UNI

Strany 36 - Appendix A: Glossary

38 •••• Appendix C: Troubleshooting KD-2300 Instruction Manual Appendix C: Troubleshooting Checklist for Basic Troubleshooting This troubleshooti

Strany 37

KD-2300 Instruction Manual Appendix C: Troubleshooting •••• 39 Problems with sensor mounting and nonrepeatable fixturing can be sources of troubl

Strany 38 - the circuit

40 •••• Appendix C: Troubleshooting KD-2300 Instruction Manual 1.CAUSE: Intermittent cable. This problem is typified by large output changes (ful

Strany 39

KD-2300 Instruction Manual Appendix C: Troubleshooting •••• 41 5.CAUSE: A malfunctioning voltmeter. SOLUTION: Check the system output voltage with

Strany 40

2 •••• Customer Service Information KD-2300 Instruction Manual Customer Service Information Should you have any questions regarding this product, pl

Strany 41 - Troubleshooting

KD-2300 Instruction Manual Service Information •••• 3 Kaman Products Standard Limited Warranty Products of Kaman Precision Products are warranted

Strany 42 - Symptoms and Causes

4 •••• Service Information KD-2300 Instruction Manual !Cautions and Safeguards This information contains important cautions and safeguards associate

Strany 43

KD-2300 Instruction Manual Service Information •••• 5 !All users, despite their level of expertise, should do the following before attempting to use

Strany 44

6 •••• Service Information KD-2300 Instruction Manual Application Information The system speeds gaging and precision measurements by eliminating er